Definition
The ICP pulse amplitude (AmpICP) is the peak-to-trough height of the cardiac pulse wave, usually about 1–4 mmHg with normal compliance. It reflects the cerebral arterial blood volume pushed in with each stroke volume, multiplied by the steepness (elastance) of the pressure–volume curve at the current ICP. Amplitude rises as compliance falls, and P2 grows until it is taller than P1. The 1–4 mmHg figure is a commonly quoted range, not a validated cutoff; amplitude has no universal abnormal threshold.
This answer was obtained on OpenEvidence using the .succinct Dotflow.
Components of the ICP waveform
- Cardiac pulse wave: P1, P2, and P3, described below.
- Respiratory wave: venous pressure changes driven by intrathoracic pressure.
- Slow waves: Lundberg A (plateau) and B waves.
The three peaks
| Peak | Name | Description and significance |
|---|---|---|
| P1 | percussive wave | Arterial systolic pulse transmitted through the choroid plexus and arteries. Its height mostly tracks arterial pressure and pulse pressure. It is normally the tallest peak. |
| P2 | tidal wave | The arterial pulse rebounding off the brain parenchyma. It reflects the elastic properties of the brain, so it is the main compliance marker. |
| P3 | dicrotic wave | Aortic valve closure. It follows the dicrotic notch. |
Normal pattern: P1 > P2 > P3.
What determines the amplitude
- Pulsatile volume: cerebral arterial stroke volume, driven by cardiac output, arterial pulse pressure, and vessel diameter. Vasodilation, for example from hypercapnia, increases it.
- Elastance: the position on the exponential pressure–volume curve. The same pulse of volume gives a bigger pressure swing where the curve is steep.
- Mean ICP: amplitude increases roughly linearly with mean ICP in TBI.
Confounders:
- Arterial pressure can change amplitude independently of compliance.
- Age: vascular and brain stiffening raises amplitude and changes waveform shape.
- Central venous pressure: a high CVP rounds the waveform.
- Technical: a damped or faulty transducer can distort the trace.
How amplitude and shape change with compliance
| Compensatory reserve | Amplitude | RAP | Waveform shape |
|---|---|---|---|
| Good (flat part of the curve) | Small; barely changes as ICP rises | Near 0 | P1 dominant |
| Shrinking (steep part of the curve) | Rises in step with ICP | Approaching +1 | P2 ≥ P1, then rounded, then triangular or pyramidal with a single peak |
| Critically high ICP (above the upper breakpoint) | Falls, even as ICP keeps rising | Negative | Sign of cerebral vessel collapse |
At critically high ICP, falling amplitude and a negative RAP mean the cerebral vessels are collapsing, autoregulation is lost, and herniation is imminent.
Clinical use
- A P2 > P1 waveform can signal poor compliance before mean ICP crosses 22 mmHg. Pathological shapes predict worse outcome even when ICP is low.
- High amplitude and a high pulse shape index (PSI) are linked to higher mortality.
- Treat RAP as a qualitative index, not a precise number.
Full English text (OpenEvidence)
Components of the ICP waveform
Cardiac pulse wave: P1, P2 and P3, described below. [5]
Respiratory wave: venous pressure changes driven by intrathoracic pressure. [6-7]
Slow waves: Lundberg A (plateau) and B waves. [5]
The three peaks
P1 (percussion wave): arterial systolic pulse transmitted through the choroid plexus and arteries. Its height mostly tracks arterial pressure and pulse pressure. Normally it is the tallest peak. [8-10]
P2 (tidal wave): the arterial pulse rebounding off brain parenchyma. It reflects the brain’s elastic properties, so it is the main compliance marker. [8-9]
P3 (dicrotic wave): aortic valve closure. It follows the dicrotic notch. [9-10]
Normal pattern: P1 > P2 > P3. [3-4][10]
What sets the amplitude
Pulsatile volume: cerebral arterial stroke volume, driven by cardiac output, arterial pulse pressure and vessel diameter. Vasodilation, for example from hypercapnia, increases it. [1-2]
Elastance: the position on the exponential pressure–volume curve. The same pulse of volume gives a bigger pressure swing where the curve is steep. [3][6]
Mean ICP: amplitude increases roughly linearly with mean ICP in TBI. [11]
Confounders:
Arterial pressure can change amplitude independently of compliance. [1][5]
Age: vascular and brain stiffening raises amplitude and changes waveform shape. [2]
Central venous pressure: a high CVP rounds the waveform. [6]
Technical: a damped or faulty transducer can distort the trace.
How amplitude and shape change with compliance
Good compensatory reserve (flat part of the curve): small amplitude that barely changes as ICP rises, with RAP near 0. Waveform shape is P1 dominant. [1][9]
Shrinking reserve (steep part of the curve): amplitude rises in step with ICP, with RAP approaching +1. Waveform shape is P2 ≥ P1, then rounded, then triangular or pyramidal with a single peak. [1][3][12]
Critically high ICP (above the upper breakpoint): amplitude falls even as ICP keeps rising, and RAP turns negative. This means the cerebral vessels are collapsing, autoregulation is lost and herniation is imminent. [1][9][13]
Clinical use:
A P2 > P1 waveform can signal poor compliance before mean ICP crosses 22 mmHg. Pathological shapes predict worse outcome even when ICP is low. [11][14]
High amplitude and high pulse shape index (PSI) are linked to higher mortality. [11]
Treat RAP as a qualitative index, not a precise number. [1]
The figure from Bögli et al. shows the pressure–volume curve with RAP zones and the matching waveform shapes:
Figure 5. Cerebrospinal pressure volume curve and pulse shape index.
Unlocking the potential of high-resolution multimodality neuromonitoring for traumatic brain injury management: lessons and insights from cases, events, and patterns. Crit Care. 31 March 2025.
Used under license from Springer Nature
Would you like to explore how waveform-derived compliance markers (P2/P1, RAP, PSI) should change TBI management when mean ICP is still under threshold?
Follow-Up Questions
Lundberg A and B wave management
Noninvasive ICP waveform monitoring accuracy
RAP-guided optimal ICP targets
Read aloud: 3 minutes
References
- Intracranial compliance monitoring using pulse shape index in traumatic brain injury: relation to cerebral physiology and clinical outcome. Critical Care. 2026. Svedung Wettervik T, Beqiri E, Hånell A, et al. Recent
- Impact of age and mean intracranial pressure on the morphology of intracranial pressure waveform and its association with mortality in traumatic brain injury. Critical Care. 2025. Kasprowicz M, Mataczyński C, Uryga A, et al.
- Intracranial pulse pressure waveform analysis using the higher harmonics centroid. Acta Neurochirurgica. 2021. Zakrzewska AP, Placek MM, Czosnyka M, Kasprowicz M, Lang EW.
- Observational study of intracranial compliance analysis in neurologically healthy pediatric patients using a non-invasive device. Scientific Reports. 2024. Karuta SCV, Folchini CM, Fachi MM, et al. Observational
- Analysis of Intracranial Pressure. The Neuroscientist : A Review Journal Bringing Neurobiology, Neurology and Psychiatry. 2013. Di Ieva A, Schmitz EM, Cusimano MD. Review
- Cerebrospinal fluid dynamics in idiopathic intracranial hypertension: a literature review and validation of contemporary findings. Acta Neurochirurgica. 2021. Kaipainen AL, Martoma E, Puustinen T, et al. Review
- The best marker for guiding the clinical management of patients with raised intracranial pressure—the RAP index or the mean pulse amplitude? Acta Neurochirurgica. 2016. Hall A, O’Kane R. Review
- Noninvasive detection of elevated ICP using spontaneous tympanic membrane pulsation. Scientific Reports. 2021. Dhar R, Sandler RH, Manwaring K, Kostick N, Mansy HA.
- Unlocking the potential of high-resolution multimodality neuromonitoring for traumatic brain injury management: lessons and insights from cases, events, and patterns. Critical Care. 2025. Bögli SY, Beqiri E, Olakorede I, et al. Review
- Noninvasive intracranial pressure monitoring in women with migraine. Scientific Reports. 2022. Rossi DM, Bevilaqua-Grossi D, Mascarenhas S, et al.
- Analysis of Intracranial Pressure Pulse Waveform in Traumatic Brain Injury Patients: A CENTER-TBI Study. Journal of Neurosurgery. 2023. Uryga A, Ziółkowski A, Kazimierska A, et al.
- Relationship between the shape of intracranial pressure pulse waveform and computed tomography characteristics in patients after traumatic brain injury. Critical Care. 2023. Kazimierska A, Uryga A, Mataczyński C, et al.
- Developmental Physiology of the Central Nervous System. Gregory’s Pediatric Anesthesia. 2020. Zoel Quinonez, R. Blaine Easley, Bruno Bissonnette, et al.
- Intracranial Pressure Monitoring in Adult Patients With Traumatic Brain Injury: Challenges and Innovations. The Lancet. Neurology. 2024. Zoerle T, Beqiri E, Åkerlund CAI, et al. Review
Follow-up questions
- Management of Lundberg A and B waves
- Accuracy of noninvasive ICP waveform monitoring
- RAP-guided optimal ICP targets
Source: OpenEvidence