The Atlas Stent Herniation Technique for Wide-Neck PCoA Aneurysms with a Fetal PCA (fPCA)

25%

PCoA aneurysms account for about 25% of all intracranial aneurysms and carry a high rupture risk — the combination of a wide neck with an fPCA is the main challenge in endovascular treatment.

The clinical problem

A PCoA (posterior communicating artery) aneurysm accounts for about 25% of all intracranial aneurysms and carries a high rupture risk. When it is wide-necked and accompanied by an fPCA, the task becomes difficult:

  • Wide neck: neck ≥ 4 mm, dome-to-neck ratio < 2.
  • fPCA (fetal posterior cerebral artery): the P1 segment is absent or hypoplastic and the PCA is effectively supplied by the ICA.
  • The aneurysm involves both the ICA and the PCoA, and both arteries must remain patent, because occluding the PCoA means losing the entire posterior territory.
  • Flow diverters also do not work well in this anatomy.

The herniation technique — concept and sequence

The Neuroform Atlas stent has a hybrid design: an open-cell body and a closed-cell proximal segment. The idea is to select a stent slightly larger than the parent vessel and deliberately allow part of it to "herniate" into the aneurysm neck — a dome that covers and protects the PCoA ostium while coiling proceeds from within the sac.

  1. 1
    The stent is partially deployedThe stent is deployed just enough to cover the aneurysm neck and form the herniation into the neck.
  2. 2
    Coiling from within the sacCoils are placed from within the aneurysm sac while the stent dome protects the PCoA ostium.
  3. 3
    Full stent deploymentAfter coiling, the stent is fully deployed. Sometimes the stent is fully deployed first and the microcatheter is passed through the stent cells — the trans-cell technique.

Sizing and contraindication

The stent diameter is selected based on the distance from the superior edge of the PCoA origin to the opposite ICA wall, not the parent-vessel diameter. The largest Atlas size is 4.5 mm, which expands to 4.9 mm.

Sizing rule
  • If the distance from the superior edge of the PCoA origin to the opposite ICA wall is more than 4.9 mm, this technique is contraindicated and one should go with Y-stenting or a flow diverter.

The study

The study was retrospective and single-center, conducted at Jingmen Central Hospital in China from 2022 to 2024, in 19 patients with a mean age of 62 years:

  • 12 with SAH (ruptured)
  • 5 incidental
  • 2 with third nerve palsy

Results

Outcome Result
Technical success 100%, with no intraprocedural complications
Immediately after the procedure Complete occlusion (Raymond I) in 13 (68.4%)
Angiographic follow-up 15 patients, median follow-up 7 months: complete occlusion in 12 (80%), with no recurrence
Vessel patency The PCoA remained patent in all patients and no in-stent restenosis was seen
Clinical outcome All 18 survivors had an mRS of 0 at final follow-up

Complications

  • One delayed thrombosis: 7 hours after the procedure, with scattered cortical infarcts in the distal MCA territory, probably due to microembolization — discharge mRS 3.
  • One death on day 3 due to multiorgan failure; the patient had remained comatose and the death was not directly related to the procedure.
  • Both complications were in the ruptured group, not the unruptured group.

Why this result matters — comparison with other options

Method Limitation / result
Y and λ (double-stent) Complex, with thromboembolism reported in up to 11%
Pipeline flow diverter in fPCA Only 1 of 9 complete occlusions; the Rinaldo study: 43.7% versus 81.8% without fPCA
WEB 54% complete occlusion
Atlas herniation (single-stent) 80% complete occlusion at follow-up — a good result with a single stent

Limitations (as stated by the article itself)

  • Small sample (19 patients), retrospective and single-center
  • Short follow-up (median 7 months)
  • Mixed ruptured/unruptured in one cohort
  • The Atlas stent is non-retrievable and invisible on angiography
  • Not suitable for aneurysms on the inner curvature, because the risk of coil protrusion rises
Key points
  • Atlas sizing in this technique is based on the distance from the PCoA origin to the opposite ICA wall, not the parent-vessel diameter; the technique’s ceiling is 4.9 mm.
  • Preserving simultaneous patency of the ICA and PCoA in the presence of an fPCA is mandatory — occluding the PCoA means losing the entire posterior territory.
  • 80% complete occlusion at follow-up (12 of 15) with a single stent, versus 54% for WEB and 43.7% for flow diverters in fPCA.

Source: Frontiers in Neurology

دیدگاه‌ خود را بنویسید

نشانی ایمیل شما منتشر نخواهد شد. بخش‌های موردنیاز علامت‌گذاری شده‌اند *

به بالا بروید